Bian Chao, Xu Yuan-yuan, Sun Hong-guang, Zhang Hong, Chen Shao-feng, Xia Shan-hong. A Micro Amperometric Immunosensor Based on MEMS[J]. Journal of Electronics & Information Technology, 2006, 28(11): 2195-2198.
Citation: Bian Chao, Xu Yuan-yuan, Sun Hong-guang, Zhang Hong, Chen Shao-feng, Xia Shan-hong. A Micro Amperometric Immunosensor Based on MEMS[J]. Journal of Electronics & Information Technology, 2006, 28(11): 2195-2198.

A Micro Amperometric Immunosensor Based on MEMS

  • Received Date: 2005-03-16
  • Rev Recd Date: 2005-07-28
  • Publish Date: 2006-11-19
  • A micro amperometric immunosensor is fabricated to detect human IgG based on MEMS. The Pt electrode is deposited on silicon wafer with the sensitive area of 1 mm2. A micro reacting pool is made of SU-8 polymer and several microlitre reagent was needed. Polypyrrole is electropolymerized on the electrode to transfer the electron between the electrode and the enzyme. Glutaraldehyde is used to bind the antibody on polypyrrole. Human IgG is specifically recognized by goat anti-human IgG and horseradish peroxidase (HRP) conjugated goat anti-human IgG in sandwich way. Human IgG is scaled by detecting the current from the catalyzed reaction between HRP and H2O2. With the working potential of #61485;-0.3V, the immunosensor displayed a good linear doseresponse behavior for human IgG concentrations between 5 and 255ng/ml.A low detection limit of 5ng/ml and 3-minute response time are achieved. This micro amperometric immunosensor has many advantages such as fast response time, low detection limit, little reagent, micro bulk and easy integration with IC.
  • loading
  • Li J, Xiao L T, Zeng G M, et al.. A renewable amperometric immunosensor for phytohormone-indole acetic acid assay[J].Analytica Chimica Acta.2003, 494(1/2):177-185[2]Santandreu M, Sold S, Fabregas E, et al.. Development of electrochemical immunosensing systems with renewable surfaces[J].Biosensors Bioelectronics.1998, 13(1):7-17[3]Darain F, Park S, Shim Y B. Disposable amperometric immunosensor system for rabbit IgG using a conducting polymer modified screen-printed electrode[J].Biosensors and Bioelectronics.2003, 18(5/6):773-780[4]Fernndez-Snchez C, Gonzlez-Garci MB, Costa-Garci A. AC voltammetric carbon paste-based enzyme immunosensors[J].Biosensors Bioelectronics.2000, 14(12):917-924[5]Campanella L, Attioli R, Colapicchioni C, Tomassetti M. New amperometric and potentiometric immunosensors for anti-humanimmunoglobulin G determinations[J].Sensors and Actuators B.1999, 55(1):23-32[6]Wilson M S, Rauh R D. Novel amperometric immunosensors based on iridium oxide matrices[J].Biosensors and Bioelectronics.2004, 19(7):693-699