He Xingjun, Wang Shouqi. STUDY OF FEASIBILITY OF Ag-AgCl ELECTRODE AS REFERENCE ELECTRODE FOR ISFET[J]. Journal of Electronics & Information Technology, 1989, 11(4): 444-448.
Citation:
He Xingjun, Wang Shouqi. STUDY OF FEASIBILITY OF Ag-AgCl ELECTRODE AS REFERENCE ELECTRODE FOR ISFET[J]. Journal of Electronics & Information Technology, 1989, 11(4): 444-448.
He Xingjun, Wang Shouqi. STUDY OF FEASIBILITY OF Ag-AgCl ELECTRODE AS REFERENCE ELECTRODE FOR ISFET[J]. Journal of Electronics & Information Technology, 1989, 11(4): 444-448.
Citation:
He Xingjun, Wang Shouqi. STUDY OF FEASIBILITY OF Ag-AgCl ELECTRODE AS REFERENCE ELECTRODE FOR ISFET[J]. Journal of Electronics & Information Technology, 1989, 11(4): 444-448.
The preparation, principle and measured results of complex ion sensitive field effect transistor (ISFET) with Ag-AgCl reference electrode are described. Through theoretical analysis and experiments, the feasibility of Ag-AgCl reference electrode which does not contact with solution and the way of overcoming its unstability are put forward. It is explained that why the effect appearing in the ISFET does not conform to Nernst effect, when Na+, Ca++ cations etc. are measured with Ag-AgCl reference electrode.
S. Shoji, M. Esashi, T. Masuo, Sensors and Actuators, S (1984), 89-99.[2]黄德培,沈子琛,关国梁等编著,离子选择电极的原理及应用,化学工业出版社,北京,1982, pp. 158-165.[3]M.惠特菲尔德著,用于天然水分析的离子选择电极,科学出版社,北京,1977, pp. 25-26.[4]谢声洛编著,离子选择电极分析技术,化学工业出版社,北京,1985, pp. 113-120.